Ie i – Heidolph Pump Heads User Manual

Page 58

Advertising
background image

I

E

I

58

-

Item's master switch (29) is located on the rear

panel. LED (40) for sense of rotation in ON.

-

Set speed with speed selector (30) on the control

panel.

-

Feed rate is proportional with pump speed. Go

by rpm scale on the item. For more information

a

chapters 16 and 17.

bout exact speed refer to reference tables in







30: Rotation speed

31: Direction of rotation

32: Start/Stop

33: Maximal

conveyance

-

Select sense of rotation with switch (31); (go by arrow marks). Sense of rotation may be

changed on standstill only.

A signal sound occurs if you press this button during operation.

Caution! Hit Stop key (32) before changing sense of rotation, select new sense of

rotation and restart.

-

Sense of rotation selected is confirmed by LEDs (40).

-

Start pump with „Start Stop" key (32). Pump status is confirmed by an LED (44).

-

To stop pump rotation, hit "Start Stop" key (32) once more.

-

When hitting "max" key (33), pump will run at max. speed as long as this key is held

depressed; this feature eases filling or draining tubing without any change to basic settings.

Bild 45

30

33

40

31

40

44

32

fig. 44

37

29

Set speed number may vary from actual number slightly.

-