9.3.1. How to program particular pump heads (figs. 49, 50)

If you have purchased a PD 5201 or PD 5206 in a product package, the drive has

already been adjusted to the respective pump head at the factory!

-

Turn power switch (29) OFF. Hit "Cal" key (34) and turn power switch (29) ON. Pump head

actually programmed is displayed.

-

Select different pump head with arrow keys (35), e.g. "Pu 2".

-

Confirm choice with "Select" key (36).

-

For coding of pump heads refer to reference table below.

code

pump head

PD 5206

PD 5201

Pu 0

multi-channel pump head C 8

X

Pu 1

SP quick

X

X

Pu 2

SP standard / SP vario

X

X

Pu 3

SP mini (obsolete)

X

X

Pu 4

Masterflex

X

X

Pu 8

multi-channel pump head C 12

X

Pu 9

multi-channel pump head C 4

X

9.3.2. Select display mode

allows to choose from several display modes:

-

speed rpm (speed in 1/min)

-

tube mm (tube ID in mm)

-

flow ml/min (feed rate in ml/min)

-

vol ml (metered volume in ml)

-

pause sec/min/h (pause in seconds, minutes or hours) (interval metering)

9.3.3. Display feed rate in ml/min

-

In order to display feed rate in ml/min, the item needs to "know" about the tubing diameter

used.

-

Feed rate is proportional with tube ID and pump speed.

-

Enter tube ID: hit

several times until

"Tube mm" is lit.

LED

-

Enter tube ID with arrow keys (35) (e.g. 0.8; 1.7; 3.1; 4.8; 6.3 or 7.9).

If the LED „Vol ml“ blinks when you adjust the internal diameter of the tube, the

metering volume is being adapted to the minimum allowable volume. (see chap.

9.3.5)

-

Hit "Select“ key (36) until LED "Flow ml/min“ is lit.

-

Feed rate in ml/min is displayed (41).

-

To alter feed rate, use arrow keys (35) after having selected the "Flow ml/min" mode with the

"Select“

key.

-

At this point, the processor is going to determine rotational speed.

"Select" key (36)

"Select“ key (36)