Ie i – Heidolph Pump Heads User Manual

Page 64

Advertising
background image

I

E

I

64

-

Volume is displayed (41) in ml.

-

Desired volume can be selected with arrow keys (35).

-

At this point the processor is going to determine the number of revolutions required.

-

Since every tube has its own tolerance, it will be required to calibrate the pump drive for a

particular tube, if high accuracy is mandatory.

9.3.6. Calibrate metered volume

-

Determine and set speed by your actual requirements

The PD 5206 drive can only meter with a maximum of 300 l/min for reasons of

precision! During the metering process, the drive is automatically limited to 300 l/min.

The LED "Speed rpm" and the display blink before the metering process as an

indication.

-

Enter tube data and volume to be metered, as described in chapter 10.3.5.

-

Hit "Start Stop" key (32).

-

Pump is going to feed fluid at the speed and volume entered before.

-

Determine exact quantity of fluid collected.

-

Hit "Cal" key (34). Display (41) flashes.

Caution: with reference to the very small quantities involved, feed rate is displayed in

µ

l/min, when

Cassette small and tube size 0.2; 0.5 or 0.9 are selected.

-

Enter value so determined with arrow keys (35).

-

Hit "Cal" key (34) and hold depressed for not less than 3 seconds to enter. Data in memory

are revised. "Cal" is displayed (41) for about 2 seconds to confirm action.

-

Enter volume wanted with arrow keys (35).

Programmed flow rate and hence volume metered are a function of pump head and

tube selected. If, during calibration, the user finds a deviation in excess of 30 %, a

setting error is very probable. In this case discontinue calibration (hit "select" key once),

and go ahead checking the following parameters:

-

did you select the right pump head (chapter 10.3.1)?

-

did you choose the right tube (chapter 10.3.3)?

-

is the tube properly installed in the pump head (chapter 3) ? Is the clamping action

ok ? (chapter 3)

If wrong parameters have been entered for calibration of metered volume, metering

action cannot be started with the "Start" key. The

"Select" will turn to "Flow ml/min",

and 4 dashes (----) are displayed.

Correct wrong parameters as described below:

-

go into feed rate graph (chapters 17 & 18) and determine a speed figure (1/min)

plus a logical feed rate (ml/min).

-

go to "speed" menu (1/min) with "Select" key and choose a speed with arrow keys.

-

go to "flow" menu (ml/min) with "Select" key and choose a feed rate from the graph,

as described in chapter 10.3.4.

-

at this point a reasonable feed rate is saved.

LED