General specifications, Source thermocouple specifications (its-90 curves) – Practical Instrument Electronics 521 User Manual

Page 3

Advertising
background image

Model 520/521

Operating Instructions

Connection Diagram

Two Wire Connection to Transmitter

Specifications

General Specifications:

(Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from
calibration)

Temperature Range

-25 to 60 °C (-10 to 140 °F)

Relative Humidity
Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size

4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)

Weight

7.2 oz (204 grams)

Battery

9V Alkaline provides 45 hours of continuous use

Miscellaneous

Low battery indication with nominal 1 hour of operation left
Overload Protected to 60V for 30 seconds or less
High contrast graphic liquid crystal display with 0.357” (9.07 mm) high digits

Source Thermocouple Specifications (ITS-90 Curves):

Millivolt Uncertainty
Temperature Coefficient of mV Source
Output Noise
Output Impedance

±(0.015% of mV Setting + 0.009 mV)
±0.005mV/°C Ambient
±5µVpp from 0.1 Hz to 10 Hz
0.2

Cold Junction Uncertainty
Cold Junction Sensor Temperature
Coefficient

±0.25°C (0.5°F)
±0.05°/° in ambient temperature (°C or °F)

General Temperature Accuracy

±(0.015% of mV setting + 0.009mV) ± 0.25°C (0.5°F)

Output Dial Adjustment Resolution

0.1°C or °F for Model 521/1°C or °F for Model 520

Span

-13.000 - 80.000 mV