Practical Instrument Electronics 522 Rev C User Manual

Thermocouple calibrator, Model 522, Operating instructions

Advertising
background image

Practical Instrument Electronics

Model 522

Thermocouple Calibrator

Operating Instructions

841 Holt Road #1 Webster, NY 14580

Tel: 585-872-9350 Fax: 585-872-2638

[email protected] http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright © 2003. All rights reserved.

522-9002 5/20/05

1-4

Basic Keypad Operations

1

EZ-Check™ Switch

SOURCE mode: Slide the switch to select from three user-stored
values for the desired calibration points. The user can select HI, DIAL
and LO positions. These values can easily be changed to suit the
calibration requirements.

READ mode: Slide the switch to recall minimum and maximum
readings. Press the EZ-Dial™ Knob to clear the stored values.

2

SOURCE/OFF/READ Switch

Slide the SOURCE/OFF/READ Switch to SOURCE
to output a voltage corresponding to the
temperature on the display for the selected
thermocouple type. Use the READ position to
directly convert thermocouple input to temperature.

3

EZ-Dial™ Knob

Turn the knob to change temperature in 0.1º
increments. Push and turn for faster dialing. Push
without turning to store new EZ-Check™ HI/LO
points in SIMULATE mode, or to clear EZ-Check™
HI/LO points in READ mode.

Model 522 Configuration

Instructions for Enabling and Disabling the Configuration Options

1. Turn the Model 522 on with the SOURCE/OFF/READ Switch.

2. Press the EZ-Dial™ Knob while the “PRESS EZ-DIAL KNOB FOR CONFIGURATION”

message is displayed.

3. Select options by turning the EZ-Dial™ Knob until the arrow points to the desired

option.

4. The option can be enabled or disabled by tapping the EZ-Dial™ Knob.

The Model 522 configuration menu will exit automatically after 5 seconds of inactivity and go
to normal operation with the options selected. These options are saved even when the unit is
turned off.