System setup, Place the hood on top of the transilluminator, High performance transilluminator – UVP DigiDoc-It Imaging System User Manual

Page 4: Benchtop transilluminator and drawer, Camera assembly

Advertising
background image

DigiDoc-It Imaging System

3

High Performance 2UV Transilluminator

(shown with UV blocking cover)

DigiDoc-It Drawer

(may be different color)

System Setup

DO NOT ATTEMPT TO CONNECT ANY WIRING WHILE THE EQUIPMENT IS CONNECTED TO ANY POWER
SUPPLY.

CAUTION: Do not install the system in places with high moisture, dust or high temperature. Do not use any oil or
petroleum based cleaner for the cabinet. Use only mild soap or detergent solution for cleaning. Ensure that the
system is turned OFF during cleaning. Keep the equipment away from motors or other large magnetic equipment
apparatus.

Place the Hood on top of the Transilluminator

The Hood Enclosure can be used with a High Performance UV Transilluminator. When using a Benchtop
Transilluminator, the optional drawer is required.

CAUTION: All UVP Transilluminators are powerful sources of UV radiation that will cause damage to unprotected
eyes and skin. Before operating any unit, be sure all personnel in the area are properly protected. UV Blocking
Eyewear is highly recommended and should be worn. Refer to the Accessories section of this manual for UV
Blocking Eyewear.

High Performance Transilluminator

1. Remove the UV blocking clear cover from the

Transilluminator.

2. Position the transilluminator so the switch is facing

the user.

3. Connect the power cord to the back of the

transilluminator and connect the plug to a power
source.

4. Place the Hood Enclosure on top of the

Transilluminator surface so that the filter glass is
covered.


Benchtop Transilluminator and Drawer

1. Remove the UV blocking clear cover from the Transilluminator.

2. Place the Benchtop Transilluminator into the drawer.

3. Connect the power cord to the back of the transilluminator and

push the plug through the access hole in the back of the
drawer.

4. Connect the plug to the power source.

5. Access to the transilluminator switch is via the access port on

the front of the drawer.

6. Place the DigiDoc-it hood over the drawer.

Camera Assembly

1. The camera will be installed in the DigiDoc-It hood at the UVP

factory.

2. To connect the power cable to the camera, thread the

camera’s power cable through the round notched hole from the
back of the camera cap and plug it into the camera as shown.