Service procedures, Return procedure, Replacement parts and accessories – UVP GelMax Imager User Manual

Page 11: Troubleshooting

Advertising
background image

GelMax Imager

11

Service Procedures

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product. See “Technical Support” below for contact information.

Replacement Parts and Accessories

To order accessories or replacement parts for the GelMax Imager, contact UVP’s offices.

Part Description

Part Number

Fuse, 2.0A

56-0022-03 (2 Recommended)

Power Cord, 100V/115V

58-0085-01

Power Cord, 230V

58-0085-03

Interchangeable Sample Plates:

365nm Longwave UV Interchangeable Sample Plate

38-0381-01

460/470nm Visi-Blue Interchangeable Sample Plate

38-0381-02

White Light Interchangeable Sample Plate

38-0381-03

Solid Black Interchangeable Sample Plate

38-0381-04

Interchangeable Sample Plate Holder

20-1657-01


Emission Filters:

Filter, Ethidium Bromide, 50mm Square

38-0220-01

Filter, SYBR Green, 50mm Square

38-0219-01

Filter, SYBR Gold, 50mm Square

38-0221-01

Additional Filters

Contact UVP

Gel Accessories:

Gel-Cutter

85-0002-01

Gel-Ruler

85-0003-01

Gel-Scooper

85-0006-01

Gel-Tray, Small

85-0007-01

Gel-Sentry DNA Preparation Plate

97-0076-01

Fluorescent Standard Step Tablet

33-0014-02

Protective Equipment:

Spectacles, UV Blocking (UVC-303)

98-0002-01

Goggles, UV Blocking (UVC-503)

98-0002-02

Faceshield, UV Blocking (UVC-803)

98-0002-04

Troubleshooting

No Power to the Transilluminator or Filter Wheel

1. Recheck the main power cord connection to the GelMax unit.

2. Check the fuses located at the back of the unit next to the power

port. A small flat-head screwdriver or similar tool will be required.

Push the left tab of the fuseholder to the right until the
fuseholder’s left side pops out. Then, push the right tab toward
the left until the right side pops out. The entire fuseholder can
now be pulled out.

Inspect the thin wire within each glass fuse to see if there is a
break in the wire. If so, replace the fuse(s). If fuses are blowing
repeatedly, contact UVP Technical Support for additional troubleshooting.