When unpacking the GelMax Imager, the following items will be included as standard:

1. GelMax Imager
2. Ethidium bromide (EtBr) emission filter
3. Power and USB cables
4. Doc-ItLS Software
5. Supporting documentation
6. One or more Interchangeable Sample Plates (the Solid Black Interchangeable Sample Plate is

included as standard)


Place the darkroom on a flat surface which can provide adequate support for up to 50 pounds.

WARNING: Do not attempt to perform any setup procedures while the system is plugged in or powered on
unless otherwise instructed.

If the equipment is used in a manner other than as specified by UVP, the protection provided by the
equipment may be impaired.

CAUTION: This system is designed for indoor use only. Do not install the system in areas with high
moisture, dust or high temperatures. Keep the equipment away from motors or any other large magnetic
equipment apparatuses.

Connecting the Power and USB Cables

Plug the square end of the USB cable into the back of the GelMax.
Plug the other end into an available USB port on the user’s computer.

Plug the main power cable into the back of the GelMax and the other
end into a surge-protected power outlet. Set the power switch on the
back of the system to the ON (I) position.

Note: It is recommended to leave the power switch on the back of the
system in the ON position except when the system will not be used for
an extended period of time (one day or longer).

Note: Do not position the system so that it is difficult to access the
power cable and operate the main power switch at the back of the unit.

Installing Emission Filters

To install the 50mm square ethidium bromide (EtBr) filter and any other emission filters:

1. Carefully remove the filter from the

protective plastic case, holding the filter
at the edges to prevent placing
fingerprints on the glass surface.

2. The filter wheel is located on the user’s

right side of the darkroom. Press in on
the slide-out filter wheel door. The door
will pop out slightly. Pull the filter wheel
assembly out until it stops.

3. Using the Emission Filter Selector Knob,

rotate the filter wheel to the desired
position. Use the chart to determine the
ideal Emission Filter Selector Knob
position for installing filters:

Slide-Out Filter

Wheel Door