Service procedures, Return procedure, Replacement parts and accessories – UVP MultiDoc-It Imaging System User Manual

Page 9: Troubleshooting, No power to the darkroom, Transilluminator will not turn on

Advertising
background image

MultiDoc-It Imaging System

9

Service Procedures

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service

before returning any product.

Replacement Parts and Accessories

Accessories are available for use with the MultiDoc-It Imaging Systems. Contact UVP’s offices at the

telephone numbers below for replacement parts for MultiDoc-It equipment. For transilluminator

replacement parts, refer to the separate transilluminator manual.

Description

Part Numbers


UV Lamps:

UVG-11, 254nm UV, 115V

95-0016-14

UVG-11, 254nm UV, 230V

95-0016-15

UVGL-15, 365/254nm, 115V

95-0017-09

UVGL-15, 365/254nm, 230V

95-0017-10

UVL-21, 365nm, 115V

95-0018-02

UVL-21, 365nm, 230V

95-0018-03

UVGL-25, 365/254nm, 115V

95-0021-12

UVGL-25, 365/254nm, 230V

95-0021-10

UV to 480nm Filter Converter

76-0310-01


Doc-ItLS Analysis Software

97-0185-02

Thermal Printer, Mitsubishi Digital (115/230V)

89-0069-06

Thermal Printer, Mitsubishi Digital (230V)

89-0069-07

Thermal Printer, Sony Digital (230V for Europe)

98-0357-01

Thermal Paper, matte 4 rolls (800 images)

89-0038-01

Thermal paper, 16 rolls (3200 images)

89-0038-04

Thermal Paper, 40 rolls (8000 images)

89-0038-05

UV

to White Light Converter Plate (21x26cm)

38-0191-01

UV to White Light Converter Plate (20x40cm)

38-0191-02

Visi-Blue Converter Plate (21x26cm)

38-0200-01

Visi-Blue Converter Plate (20x40cm)

38-0200-02

Filter, 50mm, sq., SYBR Green and EGFP

38-0219-01

Filter, 50mm, sq., Ethidium Bromide

38-0220-01

Fuse, 2 Amp, 250 V (Qty. 2)

56-0002-01

Fuse Cap (Qty. 2)

56-0036-01

Troubleshooting

No Power to the Darkroom

1. Recheck main power cord connections to the darkroom.
2. Check fuses, located at the back of the unit, near the power port. A flat-head screwdriver

will be required. Turn the fuseholder cap counterclockwise and the fuse holder will pop

out. Inspect the thin wire within the glass fuse to see if there is a break in the wire. If so,

replace the fuse(s). If fuses are blowing repeatedly, contact UVP Technical Support for

additional troubleshooting.

Transilluminator Will Not Turn On

1. There is a UV interlock switch that turns off the transilluminator and epi UV illumination

when the darkroom door is opened. Be sure the darkroom door is completely closed.

2. Be sure the darkroom cabinet’s “Power On” indicator light is lit. If not, refer to “No Power

to the Darkroom” above.