This guide can be used to prepare and run a full 1 x 96-well assay plate.
For more information on a given step, refer to the complete instruction
manual. New users can download the manual, which includes detailed
instructions and a list of kit components, at www.bio-rad.com/bio-plex.

Initial Preparation

1. Plan the plate layout.

2. Start up/warm up the Bio-Plex

®

system (30 min).

n

Bring the 10x wash buffer, assay buffer, and diluents to room

temperature (RT). Keep other items on ice until needed

n

Begin to thaw frozen samples

3. Prime wash station for flat bottom plate or set vacuum manifold to

–1 to –3" Hg for filter plate.

4. Calibrate the Bio-Plex system by following the prompts within the

Bio-Plex Manager

software. This can be done now or during an

assay incubation step.

5. Prepare 1x wash buffer. Mix 10x stock by inversion to ensure all salts

are in solution. Then dilute 1 part 10x wash buffer with 9 parts dH

2

0.

IMPORTANT!

Pay close attention to vortexing, shaking, and incubation instructions.

Deviation from the protocol may result in low assay signal and assay variability.


For use with

Instruction Manual #

Bio-Plex Pro Human Chemokine Assays 10031990

Bio-Plex Pro

Assays

Chemokine Quick Guide