Cooper Instruments & Systems PTT 242 Gage/Absolute Pressure Transmitter User Manual

Amplified transducer application sheet for ptt 242, Typical wiring to supply & readout, Maximum loop resistance

Advertising
background image



AMPLIFIED TRANSDUCER APPLICATION SHEET FOR PTT 242

Output Range 4-20 mA 2-WIRE

Operating Voltage Range 13-40 volts DC

Operating current max 28 mA (OVERRANGE)

Operating Temperature Range -40 to 225

°F

Reverse Circuit protected Yes


3.90" (50-1500 PSI)

3.15" (2K - 10K PSI)

1.50"

DIA.

1/4-18 NPT FEMALE
PRESSURE PORT w/1" HEX
(ADAPTERS USED FOR
MALE FITTINGS)

REMOVABLE SCREWS TO
ACCESS ZERO & SPAN
ADJUSTMENTS INSIDE CASE

or EQUIVALENT

CONN. GASKET

DIN 43650/GSA 200-4

or EQUIVALENT

ELECTRICAL CONNECTOR
DIN 43650/GSA 300N

Procedure for adjusting
Span & Zero Potentiometers:

1. Loosen screws on rear of unit.
2.Remove connector cover to
expose trim pots.
3.Adjust ZERO &SPAN using
SHUNT CAL or a pressure standard.
4.Replace cover, making sure not
to pinch any wires.
5.Re-insert the 2 screws on the rear of
the unit and align with mounting
Holes. CAUTION!!!!
Do not over tighten screws as
this may result in permanent
damage to the unit.




















Typical Wiring to Supply & Readout

+

-

READOUT

SUPPLY

+

-

+SUPPLY

+OUTPUT

SHUNT CAL 1

SHUNT CAL 2

3

8

13 18 23 28

38

33

40

0

500

1000

1500

SUPPLY VOLTAGE

MAXIMUM

LOOP

RESISTANCE

Maximum Loop Resistance


SHUNT CALIBRATION: (Not available on all Transducers)

To Enable the Shunt Calibration – Short together the SHUNT CAL 1 and SHUNT CAL 2 Terminals or wires.

CALIBRATION:

Using the data from the CERTIFICATION OF CALIBRATION:


Full scale OUTPUT = CALIBRATION FACTOR + ZERO OUTPUT

OUTPUT with shunt cal applied = SHUNT CAL OUTPUT + ZERO OUTPUT


ZERO OUTPUT = OUTPUT with zero pressure or load applied.


GROUNDING:

System is usually tied to Earth ground at the Power Supply Return.

PTT242 9/22/00

CF49