Grasp black ring retainer, gently wrap
tubing around rotor, insert retainer
directly into right pocket.

DO NOT

OVERSTRETCH TUBING.

Grasp thumb tab on valve and insert
firmly into left pocket. Press down
on thumb tab to ensure valve is fully
seated. Thumb tab should align with
slot in blue door. Blue door will not
close if tab is not fully seated.

Once the feeding set is loaded, close
the door. The pump is now ready for
normal operation.

Follow proper procedures for filling or spiking feed container. Press

power button in lower right-hand corner to turn pump on.