Error codes – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual

Page 62

Advertising
background image

58

Chapter 4 Alarms and Alerts

Welch Allyn Vital Signs Monitor 300 Series

Error Codes

Error
Code

Description

Blood Pressure

C01

NIBP cycle cancelled by operator.

C02

Unable to calibrate. Either the cuff is being moved excessively during the calibration or
the NIBP unit is not working.

C03

Inflation too rapid. Check for hose constriction. Verify patient type setting.

C04

Inflation too slow. Check for hose leak. Verify patient type setting.

C05

Excessive noise or air leak. Check patient condition, cuff placement and connection. Limit
patient motion.

C06

Pressure measurement outside monitor range.

C10

Cuff pressure was too high. Check the patient’s condition.

SpO

2

--

Sensor problem. Check patient condition and sensor position/contact.
Verify SpO

2

and pulse rate using an alternative method.

C7

The sensor is broken, missing, or incompatible. Contact customer service.

Temperature

C20

The probe is broken or missing or the probe well is missing. Contact customer service.

P

The temperature probe has poor tissue contact.

General

C13

Battery failure. Use the AC power adapter.

E20-E50 Internal failure. Contact customer service.