When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General

internal

malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02

Autozero failure: Check for air obstruction; limit patient movement.

C03

Inflation too rapid: Check for kinked hose or air obstruction.

C04

Excessive inflation time: Check for air leaks.

C05

Excessive noise: Check patient condition, cuff placement; limit patient movement.

C06

Measurement outside of device’s range: Check patient condition.

E10

Cuff overpressure: Check patient condition.

Temperature Error Codes

C20

Broken/missing probe: Replace probe.

P

Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

SpO

2

Error Codes

C6

SpO

2

heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11

Internal safety violation: Check patient; contact Technical Service.

C12

Ambient temperature out of range: Adjust room temperature or device location.

C13

Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General

internal

malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.