Clarion Platinum Sound Processor User Manual

Page 29

Advertising
background image

24

Platinum Sound Processor User Guide

Performing a System Check

You can verify CLARION’s operating status in three simple steps:
1) Check battery charge status
2) Check lock status
3) Check microphone and system status

To assure the system’s overall proper functioning, perform all three steps.
Once you get used to the steps, it takes less than one minute to complete a
CLARION system check.

1. Check Battery Charge Status

Turn the Processor Program Switch to the off (o) position and then on to
the desired program position --

!,!!, or .


Next, observe the red LED indicator. The red LED Light sequence will
indicate the battery charge status, as follows (Note: The LED Indicator
light shows the status of the three tests. Depending on the purpose of the
tests, the light is either red or green, as described in this section):

4 quick blinks indicate that the battery is fully charged.
2 to 3 quick blinks
indicate that the battery is sufficiently charged to
power the system.
1 quick blink
indicates that the battery charge is nearly depleted. The
battery should be replaced.


Instructions on how to change the battery are provided later in this section.

2. Check Lock Status


The term lock refers to the successful transmission of continuous signals
between the Headpiece and the Implant across the skin.

To check the lock status, continue to observe the LED Indicator. After the
fast blinking battery test sequence, the red LED will continue to flash

!

! !

Advertising