Manual No. M13-167-B0614

Range

Figure II-6

TIP: The Range value is the same as the AutoRange available on earlier Brookfield

viscometer models.

II.7 Out of Range

The DV2T Viscometer will give on screen indications when the measurement is out of range

of the instrument. When the %Torque reading exceeds 100% (over range), the display

of %Torque, Viscosity, and Shear Stress will be EEEE (see Figure II-7). If the %Torque

value is between 0 - 9.9%, the data field label will flash. When the %Torque is below zero

(negative values), the display of Viscosity and Shear Stress will be ----.

TIP: Brookfield recommends collecting data only when the %Torque reading is

between 10 - 100%.

Figure II-7