Brookfield DV2T Viscometer User Manual

Page 26

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 26

Manual No. M13-167-B0614

Figure III-3

Figure III-4

Single Point

: Collect only a single data point when the End Condition is met.

Single Point Averaging

: Specify an amount of time over which to average measured

data. Collect a single data point when the End Condition is

met. This data point is an averaged value. If the time for

averaging is shorter than the total time for the step, then the

average will be performed for the specified time at the end

of the test.

Example 1: The End Condition is Time with a value of 1 minute and 30 seconds,

the Single Point Averaging Duration is 30 seconds, the single data point

collected from this step will be an average of the data measured from 1

minute to 1 minute 30 seconds.

Advertising