III.8 Running a Test

A viscosity test is started by pressing the Run button on the Configure Viscosity Test

screen. When Run is pressed, the display will change to the Running Viscosity Test screen

(see Figure III-9).

Trend Bar

Command Keys

Status Bar

Title Bar

Test Name

M e a s u r e m e n t
Data

Figure III-9

The Running Viscosity Test screen provides information on the current measurement

including: Torque, Viscosity, Shear Stress, Shear Rate, Temperature and Speed.

Torque

is the deflection of the Viscometer torque sensor. It is described as a percent

(%) and has a range of 0 – 100%. The DV2T will provide measurement results

within the stated accuracy provided the Torque reading is between 10 and 100%. If

the Torque reading falls below 10% the labels in the data fields will flash to indicate

an error condition. Brookfield does not recommend that data be collected below 10%

Torque; however, data collection is not restricted.

TIP: Torque on the DV2T is equivalent to the dial reading from the Brookfield Dial

Reading Viscometer or the % reading from Brookfield Digital Viscometers

(DV-E, DV-I, DV-II, DV-III).

Viscosity

is calculated from the measured Torque based on the selected spindle and

speed of rotation. The units of viscosity are defined in the Global Settings section of

the Settings Menu

. If the Torque reading falls below 10%, the labels in the data

fields will flash to indicate an error condition.

Shear Stress

is calculated from the measured Torque based on the selected spindle.

The units of shear stress are defined in the Global Settings section of the Settings
Menu

. If the Torque reading falls below 10%, the labels in the data fields will flash

to indicate an error condition.

TIP: Shear Stress will be displayed as zero for spindles that do not have SRC

values.