Brookfield DV2T Viscometer User Manual

Page 36

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 36

Manual No. M13-167-B0614

Post Test Averaging

: Average and Standard Deviation are calculated for measured and

calculated parameters including: Viscosity, Torque, Shear Stress,

and Temperature.

Post Test Averaging is calculated regardless of the Data Collection

setting. If Data Collection was set to Multipoint Averaging, then the

Post Test Averaging will calculate an average of the averaged data.
When using multiple step programs generated through PG Flash

software, the test step utilized for Post Test Averaging is specified

in the Report setting (see Section III.7).

Test Used

: Display the test elements used to generate the data set. In this view, the

Configure Test button is available. Selecting Configure Test will program the

DV2T to run the same test utilized to collect the data set being viewed and

present the Configure Viscosity Test screen.

Notes

: Document any relevant information about the test or data. This information will

be stored with the data set once saved.

Compare Results

: Load two data sets and view side by side.

Figure III-12

Device Info

: Display basic information about the data file and the specific DV2T Viscometer

used to collect the data including: date and time the test started, completed

and was saved; if user accounts are active, which user saved the data;

viscometer serial number, firmware version, viscometer Torque range (LV,

RV, HA, HB, other).

III.10 Data Averaging

The DV2T Viscometer offers two techniques to average data, Live Averaging and Post Test

Averaging. Data averaging can be useful when measuring samples with entrained air or

suspended particles that case some variation in measurement results. Data averaging

may not be useful when changes in measurement results are caused by the rheological

properties of the test sample such as thixotropy or pseudoplasticity (shear thinning).