IV.4.3 Set Time and Date

The Time and Date is displayed on the Status Bar at the top of every screen. These

parameters are set within the Set Time and Date menu. The format for setting Time and

Date will be based on the Regional Settings (see Section IV.3). Change from Set Time to

Set Date by pressing the command key at the bottom of the screen. When both Date and

Time are set correctly for your location, press OK to accept the change.

IV.4.4 Backup and Import

The DV2T Viscometer provides several options relating to storage of Results, Tests and

User Settings. Backup and Import menu options relate to creating back up files, updating

instrument software, and updating language files.

Backup

: Create a backup file from the DV2T internal memory including: Settings, User

Profiles, Results, and Tests. This backup file is saved to a USB Flash Drive and can be

uploaded (Restore) to any DV2T Viscometer.

Choose the location within a USB Flash Drive where the Backup file is to be saved. A

new folder can be created by using the New Folder button. Press the Backup button to

create the Backup file.

Restore

: Upload a previously saved Backup file. The information from the Backup file

includes: Settings, User Profiles, Results, and Tests. Existing User profiles will not be

overwritten. User profiles from the Backup file will be added to the user profiles that already

exist on the DV2T.

Locate the Backup file on the attached USB Flash Drive. Press the Restore button to

upload the Backup file.

Automatic Backup

: Create Backup files according to a schedule. The frequency is set to

Daily, Weekly or Monthly increments. The Path must be specified and requires the use of

a USB Flash Drive. Check the Allow Automatic Backup check box to initiate this feature.

The current Backup file will overwrite the previous Backup file located in the path location.

TIP: You can preserve historical backup files by using multiple USB Flash Drives.

Import

: Upload Settings from a Backup file. User profiles, Results and Tests are not

included.

Locate the Backup file on the USB Flash Drive. Press the Import button to upload Settings

from the Backup file.

Add Language

: Language files can be added to the DV2T as they become available

from Brookfield. Insert a USB Flash Drive into the DV2T that contains the language file.

Select Add Language from the Backup and Import menu and select the appropriate path.

Press the Import command button to bring the new language file into the DV2T. Select

the language in the Display section of the User Settings menu.

Update Software

: The internal operating software of the Brookfield DV2T can be updated

from a USB Flash Drive as new versions become available from Brookfield. Insert a USB

Flash Drive into the DV2T that contains the operating software file. Select Update Software

from the Backup and Import menu and select the appropriate path. Press the Update

command button to bring the new software file into the DV2T.