Brookfield DV2T Viscometer User Manual

Page 54

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 54

Manual No. M13-167-B0614

IV.4.8 Calibration Reminder

The DV2T can be programmed to provide the user with a reminder that calibration is due.

Two parameters must be set: Frequency in Months and Start Date. When both parameters

are entered, the Calibration Reminder screen (see Figure IV-11) will calculate and display

the day when the reminder will be presented.
Check the Calibration Reminder On check box to activate this feature.

Figure IV-11

The Calibration Reminder will be presented during the first power up on or after the specified

day. The message box (see Figure IV-12) requires that the user press OK before the

AutoZero will be performed.

Figure IV-12

The calibration reminder will be reset by Brookfield or your authorized Brookfield dealer

when the unit is returned for calibration.

IV.4.9 Save Audit Trail

The Audit Trail is a collection of instrument check values that can assist Brookfield with

troubleshooting activities. You may be asked to provide the Audit Trail files. Use Save

Audit Trail to create an Audit Trail file on a USB stick. Two files will be created: Users.bin

and Service.bin.