Brookfield DV2T Viscometer User Manual

Page 56

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 56

Manual No. M13-167-B0614

Spindle: The spindle to be used with this

Viscosity Test. All calculations will be

based on this spindle number.

Report Screen: Select the view to be presented upon

the conclusion of the Viscosity Test.

Save Results Path: Select if a path for saving data is to be

defined. Define the save path.

Post Test Averaging: Select the type of averaging to perform

at the end of the test.

Instructions: Document the instructions to present

to the operator when the test is run.

The Test Parameters section includes all Viscosity Test program controls that may vary

if there is more than one test step.

Speed: Speed of rotation of the spindle.

Data Collection: Data Collection method.

Time Interval: Time interval used for Data Collection

methods that include averaging.

Density: Density value to be used if kinematic

viscosity units are selected (cSt, mm2/s).

End Condition: Condition which sets the end of the

Viscosity Test.

End Condition Value: Parameter based on chosen End

Condition.

QC Limits: Parameter to be used for QC limit

indicators.

QC High / QC Low Limit Parameter based on the chosen QC

Limits.

Collect Single Point at Step End: During a Multi Point data collection,

force the DV2T to collect a data point

at the end of the step (when the End

Condition is achieved).

Include This Step into Average

Calculation:

Data from this step will be included in

the average calculation indicated by the

selection in Post Test Averaging.

Add, Insert, Delete Step Buttons

Save Button

The Program Grid shows all Test Parameters for each step in a line view.

All measurement units can be set through the Units Setup button, see Figure V-2. This

menu can be pinned to the screen. If the menu is not pinned, then it will minimize when

the mouse is moved outside of the window.