Brookfield DV2T Viscometer User Manual

Page 61

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 61

Manual No. M13-167-B0614

A.4 Verifying Calibration

1. Determine the appropriate sample volume. Refer to Table A-1 to determine the correct

sample volume required for the spindle to be utilized.

2. Select a Brookfield Viscosity Standard fluid that will give viscosity readings between

10% and 100% of full scale range. Refer to Appendix B for viscosity ranges of cone

spindles.

Brookfield uses mineral oil viscosity standard fluids to calibrate Wells Brookfield Cone/

Plate Viscometers at the factory. Brookfield recommends that customers use mineral

oil viscosity standard fluids when you perform a calibration check.

If you decide to use a silicone viscosity standard fluid, do not use a fluid with a viscosity

value greater than 5000 cP with a Cone/Plate. Brookfield offers a complete range of

mineral oil viscosity standards suitable for use with Cone/Plates for viscosities above

5,000 cP or shear rates above 500 sec

-1

; see Table F-2 in Appendix F for a list of available

fluids.

It is best to use a viscosity standard fluid that will be close to the maximum viscosity for

a given cone spindle/speed combination.

Example

: DV2TLV Viscometer, Cone Spindle CPA-42Z, Brookfield Silicone Viscosity

Standard having a viscosity of 9.7 cP at 25°C

At 60 RPM, the full scale viscosity range is 10.0 cP. Thus, the Viscometer reading

should be 97% torque and 9.7 cP viscosity ± 0.197 cP. The allowable error (±0.197 cP)

is a combination of Viscometer accuracy and fluid tolerance (refer to Interpretation of

Calibration Test Results

in Appendix F).

3. With the motor off, remove the sample cup and place the viscosity standard fluid into

the cup.

Table A-1

Cone Part No.

CPA-40Z, CPE-40, CP-40

CPA-41Z, CPE-41, CP-41

CPA-42Z, CPE-42, CP-42

CPA-51Z, CPE-51, CP-51

CPA-52Z, CPE-52, CP-52

Sample Volume

0.5 mL
2.0 mL
1.0 mL
0.5 mL
0.5 mL

4. Attach the sample cup to the Viscometer and allow sufficient time for the sample, cup

and cone to reach temperature equilibrium.

5. Turn the motor on. Set the desired speed(s). Measure the viscosity and record the

reading in both % torque and centipoise (cP).

6. Verify that the viscosity reading is within the allowable 1% deviation, as explained earlier,

for the specific viscosity standard fluid(s) that you are using.

The CPE or CPA designation on the cone spindle indicates use with Electronic Gap

Setting Cone/Plate Viscometers/Rheometers only.

Advertising