Appendix F - Calibration Procedures

The accuracy of the DV2T is verified using viscosity standard fluids which are available from

Brookfield Engineering Laboratories or your local Brookfield agent. Viscosity standards are

Newtonian, and therefore, have the same viscosity regardless of spindle speed (or shear rate).

Viscosity standards, calibrated at 25°C, are shown in Table F-1 (Silicone Oils) and Table F-2

(Mineral Oils).

For more help, you can go to the website,

www.brookfieldengineering.com, and download the video.


Container size

: For Viscosity Standards < 30,000 cP, use a 600 mL Low Form Griffin

Beaker having a working volume of 500 mL.

For Viscosity Standards

30,000 cP, use the fluid container.

Inside Diameter: 3.25”(8.25 cm)

Height:

4.75”(12.1 cm)

Note: Container may be larger, but may not be smaller.

Temperature

: As stated on the fluid standard label: (

+

/-) 0.1

°

C

Conditions

:

The DV2T should be set according to the operating instructions. The

water bath must be stabilized at test temperature. Viscometers with the

letters “LV” or “RV” in the model designation must have the guard leg

attached (see Appendix G for more information on the guard leg).

Normal 25ºC Standard Fluids

Viscosity (cP)

Viscosity (cP)

High Temperature Standard Fluids

Three Viscosity/Temperatures**

5

5,000

HT-30,000

10

12,500

HT-60,000

50

30,000

HT-100,000

100

60,000

500

100,000

**25

°C, 93.3°C, 149°C

1,000

Refer to Brookfield catalog for more information

Table F-1

MINERAL OIL VISCOSITY STANDARD FLUIDS

BEL Part No.

Viscosity (cP) 25

°C

B29

29

B200

200

B600

600

B1060

1,060

B2000

2,000

B10200

10,200

B21000

21,000

B73000

73,000

B200000

200,000

B360000

360,000

Table F-2