Brookfield DV2T Viscometer User Manual

Page 8

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 8

Manual No. M13-167-B0614

I.2 Utilities

Input Voltage:

Universal Power Supply (90-264 VAC)

Input Frequency:

50/60 Hz

Power Consumption:

150 VA

Power Cord Color Code:

United States

Outside United States

Hot (live)

Black

Brown

Neutral

White

Blue

Ground (earth)

Green

Green/Yellow

Main supply voltage fluctuations are not to exceed ±10% of the nominal

supply voltage.

I.3 Specifications

Speeds:

0.1 - 200 RPM

Weight:

Gross Weight: 23 lbs. 10.5 kg.

Net Weight:

20 lbs. 9 kg.

Carton Volume: 1.65 cu. ft.

0.05 m

3

Carton Dimensions: 22 in. (56 cm) W x 11 in. (28 cm) L x 22 in. (56 cm) H

Temperature Sensing Range:

-100˚C to 300˚C (-148˚F to 572˚F)

USB A Port for use with an attached PC.

USB B Port (x3) for use with a USB Flash Drive.

Viscosity Accuracy:

±1.0% of full scale range

The use of accessory items will have an effect on the

measurement accuracy. See Appendix B.

Viscosity Repeatability:

±0.2% of Full Scale Range

Temperature Accuracy:

±1°C | -100°C to +149°C

±2°C | +150°C to +300°C

Operating Environment:

0°C to 40°C temperature range (32°F to 104°F)

20% - 80%R.H.: non-condensing atmosphere

Ball Bearing Option:

If you ordered the ball bearing suspension system with your new instrument please

note the following:

1) The ball bearing suspension in your Brookfield instrument is noted on the serial

tag on the back of the head by the letter “B” in the part number (the ninth digit;

for example: XDV2THBTB00U00).

2) When attaching and detaching the spindle, it is not necessary to lift the coupling

where the spindle connects to the instrument.

3) The Oscillation Check explained in Section IV.1: Device Setup, does not pertain

to this instrument.