More Solutions to Sticky Problems. It is important to note that for many viscometer users

the true viscosity is not as important as a repeatable day to day value. This repeatable value

can be obtained without any special effort for any measurement circumstance. But, it should

be known that this type of torque reading will not convert into a correct centipoise value when

using a Brookfield factor if the boundary conditions are not those specified by Brookfield.

The guard leg is a part of the calibration check of the Brookfield LV and RV series Viscometer/

Rheometer. Our customers should be aware of its existence, its purpose and the effect that

it may have on data. With this knowledge, the viscometer user may make modifications to

the recommended method of operation to suit their needs.

4 11/16

1 7/16

3 3/16

5 9/32

B-21Y

Guard Leg

For RV Torque

B-20Y

Guard Leg

For LV Torque

Figure G-1 - Brookfield Guard Leg

B-21KY

Guard Leg

For RV Torque

B-20KY

Guard Leg

For LV Torque

5 7/8

5 5/32

3 3/16

1 7/16

Figure G-2 - Brookfield EZ-Lock Guard Leg