A&D Weighing SCE-03 RS-232C User Manual

A&D Weighing Hardware

Advertising
background image

4

1

Print switch mode / Command mode Function

setting

prt 2

When

the

PRINT

switch is pressed while the weight value is stable (STABLE indicator is on), the

scale transmits the value.

Auto-print mode +/- data / Command mode Function

setting

prt 3

The scale transmits the weight value when the display is stable (STABLE indicator is on) and the value

is more than +4d or less than -4d. The next output can be obtained after the value returns to be more
than -4d and less than +4d.
d = minimum display (Refer to “

Specifications” of the scale instruction manual.)

Even the counting mode uses “d” for judgment.

Auto-print mode + data / Command mode Function

setting

prt 4

The scale transmits the weight value when the display is stable (STABLE indicator is on) and the value

is more than +4d. The next output can be obtained after the value returns to be less than +4d.
d = minimum display (Refer to “

Specifications” of the scale instruction manual.)

Even the counting mode uses “d” for judgment.


3.4. Command mode

In the command mode, the scale is controlled by commands that come from the external device such
as a personal computer.

Command list

• “ Q ” command

Command to request the current weight value.

Command

Q C

R

L

F

Reply

S T , + 0 0 1 2 3 . 4 5 _ k g C

R

L

F

• “ Z ” command

Same operation as the

ZERO

switch.

Command

Z C

R

L

F

• “ T ” command

Same operation as the

TARE

switch.

Command

T C

R

L

F


Reply to the command

• When the command cannot be executed, for example, because the scale is unstable, “ I ” will be sent.

Reply

I C

R

L

F

• If the received command is not for the SC/SE series scale, the scale will send “ ? ”.

Reply

? C

R

L

F

• When “aCk 0” is selected, there is no reply except the “ Q ” command.

SCE-03 RS-232C Serial Interface and

Comparator Relay Output

Instruction Manual

1WMPD4002182A

1. Features

• This interface allows the scale to be connected to a printer or a personal computer.
• It also allows outputting the comparison results of

HI

,

OK

or

LO

obtained using the comparator

function as a relay signal.

Note: SCE-02 and SCE-03 can not be installed at the same time.


2. Installation Procedure

Note: Before installing the SCE-03 board, be sure to remove

the batteries from the scale.

Step 1 Pass the cable to connect an external device through the hole

on the casing and connect it to the terminal block on the
interface board.

Step 2 Insert the interface board into the casing.
Step 3 Place the waterproof packing on the casing, aligning its holes

with the casing screw holes.

Step 4 Remove the four screws on the option slot panel located on

the bottom of the display.

Step 5 Insert the interface board into the option slot.
Step 6 Secure the casing to the display, using the four screws

provided with the casing.

Step 7 Seal the screw heads with the tabs attached to the waterproof

packing.

Cable connection using optional cables (sold separately)
When connecting to an external device with hardware flow

control, communication will be impossible using a cable
without RTS and CTS connected. In that case, connect RTS
and CTS. This will disable hardware flow control, but enable
communication.
• When the AX-KO3285-320 cable is used, RTS and CTS are

internally connected and the above operation id unnecessary.

• When the connector of an external device is a D-sub 9-pin

connector, pin 7 is RTS and pin 8 is CTS.

AX-KO3285-320





3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN

Telephone: [81] (3) 5391-6132

Fax: [81] (3) 5391-6148

Cable (AWG 26~16) to

connect an external device

Waterproof packing

Seal the screw heads

Casing

Screws on the

option slot

6 mm (AWG 26~16)

Terminal block

FG

SG

DSR

TXD

RXD

COM

LO

OK

HI

The hole for

the cable

Interface board

Cable (AWG 26~16) to

connect an external device

The hole for

the cable

1. Brown (DCD)

2. Red (RXD)

3. Black (TXD)

4. Yellow (DTR)

5. Green (SG)

6. Blue (DSR)

Shield

RXD

TXD

DSR

SG

FG

SCE-03

Terminal block J1

Optional cable unit to

connect with a PC (3 m)