American Diagnostic Corporation (ADC) Sphygmomanometer User Manual

Sphygmomanometer accessories, Warranty, Directions for use, care, & maintenance

Advertising
background image

ADC

55 Commerce Drive

Hauppauge, NY 11788

U.S.A.

ADC (UK) Ltd.

Unit 6, PO14 1TH

United Kingdom

Inspected & Packaged in the U.S.A.

tel: 631-273-9600

toll free: 1-800-232-2670

fax: 631-273-9659

www.adctoday.com

[email protected]

ADC

®

Sphygmomanometer

Accessories

Directions

for Use, Care, &

Maintenance

Warranty

American Diagnostic Corporation’s (ADC

®

) warranty service extends to the original

retail purchaser only and commences from the date of delivery. ADC warrants its

products against defects in materials and workmanship under normal use and service

as follows:

• Your Manometer is warranted for 10 years, 20 years, or Life depending upon

the model. Refer to label or end panel for specific warranty denotation.

The manometer is warranted to remain accurate to +/-3mmHg (or the

prevailing standard) over its full range when compared to a reference standard

for life.

• Adcuff

TM

inflation system components (cuff, bladder, tubing, bulb, valves,

connectors) are warranted for three years. Specialty cuffs and bladders are

warranted for 2 years.

What is Covered: Replacement of parts, and labor.

What is Not Covered: Transportation charges to and from ADC

®

. Damages

caused by abuse, misuse, accident, or negligence. Incidental, special, or

consequential damages. Some states do not allow the exclusion or limitation of

incidental, special, or consequential damages, so this limitation may not apply

to you.

To Obtain Warranty Service: Send item(s) postage paid to ADC

®

, Attn: Repair

Dept., 55 Commerce Dr., Hauppauge, NY 11788. Please include your name and

address, phone no., proof of purchase, and a brief note explaining the problem.

Implied Warranty: Any implied warranty shall be limited in duration to the terms

of this warranty and in no case beyond the original selling price (except where

prohibited by law). This warranty gives you specific legal rights and you may have

other rights which vary from state to state.

IB p/n 9355N-00 rev 2

Printed in the U.S.A.

®

AMERICAN DIAGNOSTIC CORPORATION

Maintenance:

Clean and disinfect as required

Cleaning

• Use one or more of the following methods and allow to air dry:

• Wipe with mild detergent and water solution (1:9 solution). Rinse.

• Wipe with Enzol per manufacturer’s instructions. Rinse.

• Wipe with .5% bleach and water solution. Rinse.

• Wipe with 70% isopropyl alcohol.

• Launder with mild detergent in warm water, normal wash cycle.

Remove bladder first. Cuff is compatible with 5 wash cycles.

Low Level Disinfection

Prepare Enzol enzymatic detergent according to the manufacturer’s instructions.

Spray detergent solution liberally onto cuff and use a sterile brush to agitate the

detergent solution over entire cuff surface for five minutes. Rinse continuously

with distilled water for five minutes. To disinfect, first follow the cleaning steps

above, then spray cuff with 10% bleach solution until saturated, agitate with a

sterile brush over entire cuff surface for five minutes. Rinse continuously with

distilled water for five minutes. Wipe off excess water with sterile cloth and allow

cuff to air dry.

WARNING: Do not allow a blood pressure cuff to remain on patient for more

than 10 minutes when inflated above 10 mmHg. This may cause patient

distress, disturb blood circulation, and contribute to the injury of

peripheral nerves.

WARNING: Do not apply cuff to delicate or damaged skin. Check cuff site

frequently for irritation.

WARNING: Safety and effectiveness with neonate cuff sizes 1 through 5 is

not established.

WARNING: Only use the cuff when the range markings indicated on the cuff

show that the proper cuff size is selected, otherwise erroneous readings

may result.

WARNING: Allow space between patient and cuff. Two fingers should fit in

this space if the cuff is correctly positioned.

WARNING: Do not apply cuff to limbs used for IV infusion.

WARNING: Patient should remain still during measurement to avoid

erroneous readings.

CAUTION: Do not iron cuff.

CAUTION: Do not heat or steam sterilize cuff.

Bulb and Valve Assembly - Model 872N

Intended Use

ADC

®

blood pressure bulb and valves allow for

controlled inflation and deflation of the blood pressure

cuff. They are intended for use with manual

noninvasive blood pressure measurement devices.

To Operate:

Close the valve by fully rotating thumbscrew clockwise. To open, rotate

counterclockwise. Deflation should be maintained at a consistent rate of

2-3mmHg/second throughout measurement for best results (in accordance

with recommendations set forth by the AHA).

Maintenance:

Your bulb and valve assembly is equipped with an end valve that allows the intake

of air into your blood pressure device. It is recommended that the end valve be

checked for debris and cleared on a regular basis to ensure that dust and other

debris does not reduce airflow. NOTE: Some end valves are equipped with a filter

screen to help prevent the ingress of dust from the environment.

Manometer Quality Control

A Serial number and Lot number are automatically

assigned to every aneroid during manufacturing,

ensuring every item is "controlled".

This Serial Number can be located on the

faceplate of each aneroid (Figure 6).

The Lot number is located on the outside

label of all accessory packaging

(Figure 7).

Disposal

When your sphygmomanometer or any of its parts have reached its end of life,

please be sure to dispose of it in accordance with all regional and national

environmental regulations. Devices that have become contaminated should be

disposed of in accordance with all local ordinances and regulations.

Symbol Definitions

The following symbols are associated with your ADC

®

Sphygmomanometer

!

!

!

!

!

!

!

!

!

(Figure 6)

Serial Number

This manual is available online in a variety of

languages, follow the links for language options.

www.adctoday.com/care

To register your product visit us at

www.adctoday.com/register

FOR QUESTIONS, COMMENTS,

OR SUGGESTIONS CALL TOLL FREE:

1-800-ADC-2670

OR VISIT

www.adctoday.com/feedback

LOT

: 0000

LOT 0 00000 000

(Figure 7)

Lot Number

Symbol

Definition

Important Warning/Caution

Latex-Free

Circumference Size

Conforms to EU Standards

Symbol

Definition

Authorized European
Represenative’s Information

Manufacturer’s Information

Temperature Limit

Humidity Limitation

°

F