6 reset function – KERN & SOHN EMB 5.2K5 User Manual

Page 15

Advertising
background image

6.2.6 Reset function

Factory setting reset.

 When the balance is switched off, press the OFF -key and keep it pressed

 Do not release the OFF -key. Press also the ON/TARE-key and keep it pressed

 Keep the ON/TARE-key pressed, however release the OFF -key

 Press OFF -key anew

 Keep both keys pressed until in the display “tr” appears

 Release both keys. The balance is situated in the menu.

 Press the ON/TARE-key until „rSt“ appears

 Confirm with the OFF key

 Press the ON/TAREkey to switch among the settings:

Display

Mode

„rst“ YES

Factory setting reset

„rst“ NO

No factory setting reset

 Confirm with the OFF key. Wait until the balance automatically returns to weigh-

ing mode

EMB-BA-e-1434

15

Advertising