5.8 Adjustment

With an adjustment weight, the weighing accuracy can be checked and re-adjusted at any
time.

Attention: In the verified balances the adjustment is not possible.

Procedure when adjusting:
Observe stable environmental conditions. A short warming up time of ca. 15 minutes is
recommended for stabilization.

5.9 Verification

General introduction:

According to EU directive 90/384/EEC balances must be verified if they are used as follows
(legally controlled area):

a) For commercial transactions if the price of goods is determined by weighing.

b) For the production of medicines in pharmacies as well as for analyses in the medical

and pharmaceutical laboratory.

c) For official purposes

d) For manufacturing final packages

In cases of doubt, please contact your local trade in standard.

Verification instructions

An EU type approval exists for balances described in their technical data as verifiable. If a
balance is used where obligation to verify exists as described above, it must verified and re-
verified in regular intervals.
Re-verification of a balance is carried out according to the respective national regulations.
The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!