6 connecting to the weighing terminal, 4 operating limits – KERN & SOHN IXS 300K-2M Platform User Manual

Page 6

Advertising
background image

KXP_V20-BA-e-1310

6

3.6 Connecting to the weighing terminal

Deadweight cell output

KERN KXP V20 weighing platform connection

EXC+

See marking of the deadweight cell

EXC-
SIG-
SIG+

4 Operating limits

The weighing platform is designed so robustly that an occasional exceeding of the

maximum weighing load does not lead to damage.

The static bearing capacity, i.e. the maximum permissible load, depends on the type

of load carrying (position A – C). The maximum static bearing capacity may not be

exceeded.

Avoid falling loads, shock loads as well as im-

pacts from the side.

A

at centered load

B

at load on side

C

at one side corner load


Modell

A

B

C

KXP 6V20LM

9 kg

6 kg

3 kg

KXP 15V20M / KXP15V20LM

22 kg

15 kg

7 kg

KXP 30V20M / KXP30V20LM

45 kg

30 kg

15 kg

KXP 60V20M / KXP 60V20LM

90 kg

60 kg

30 kg

KXP150V20M / KXP 150V20LM 225 kg

150 kg

75 kg

KXP300V20M

450 kg

300 kg

150 kg

Advertising