3 preload, deadload and overload settings – KERN & SOHN IXS 300K-2M Platform User Manual
Page 9
9
KFP_V20≤300-BA-e-1218
6.3 Preload, Deadload and Overload settings
Kern model
max. Preload* (kg)
* = additional initial
load
Deadload** (kg)
**= initial load placed
earlier
Center Overload
Protection circa (kg)
Corner Overload
Protection circa (kg)
Loadcell
Capacity (kg)
KXP 6V20 LM
1.86
2.14
8.5
4.8
10
KXP 15V20 M
2.86
2.14
23
12
30
KXP 15V20 LM
2.86
4.48
23
12
30
KXP 30V20 M
10.52
4.48
46
24
50
KXP 30V20 LM
10.52
9.02
46
24
50
KXP 60V20 M
35.52
4.48
85
48
100
KXP 60V20 LM
35.52
9.02
85
48
100
KXP 150V20 M
90.98
9.02
200
120
200
KXP 150V20 LM
136.14
13.86
270
120
300
KXP 300V20 M
186.14
13.86
550
240
500
Platform type
Platform
dimension
(mm)
Load-
cell
TC
Class
Max
E
max
E
min
Y
V
min
n
Dead-
load
T
min
T
max
Z
Cable- P
Lc
Typ
Nr.
Preload
-1
-4
-2
-3
(kg)
-5
-6
oder length
(kg)
(kg)
(g)
(g)
DR
(m)
KXP 6V20 LM
300x240x86
L6D
D09-03.20
C3
1.86
10
0
5000
2
3000
2.14
-10
40
n
LC
3
0,7
KXP 15V20 M
300x240x86
L6D
D09-03.20
C3
2.86
30
0
5000
2
3000
2.14
-10
40
n
LC
3
0,7
KXP 15V20 LM
400x300x89
L6D
D09-03.20
C3
10.52
30
0
5000
10 3000
4.48
-10
40
n
LC
3
0,7
KXP 30V20 M
400x300x89
L6E
D09-03.21
C3
10.52
50
0
6000
10 3000
4.48
-10
40
n
LC
3
0,7
KXP 30V20 LM
500x400x130
L6E
D09-03.21
C3
10.52
50
0
6000
20 3000
9.02
-10
40
n
LC
3
0,7
KXP 60V20 M
400x300x89
L6G
D09-03.22
C3
35.52
100
0
6000
20 3000
4.48
-10
40
n
LC
3
0,7
KXP 60V20 LM
500x400x130
L6G
D09-03.22
C3
35.52
100
0
6000
50 3000
9.02
-10
40
n
LC
3
0,7
KXP 150V20 M
500x400x130
L6G
D09-03.22
C3
90.98
200
0
6000
50 3000
9.02
-10
40
n
LC
3
0,7
KXP 150V20 LM
650x500x132
L6G
D09-03.22
C3
136.14
300
0
6000 100 3000
13.86
-10
40
n
LC
3
0,7
KXP 300V20 M
650x500x132
L6G
D09-03.22
C3
186.14
500
0
6000 100 3000
13.86
-10
40
n
LC
3
0,7