Cleaning and maintenance

Unplug the mains plug from the power unit IEX-A01.

Check, whether the electrostatic earth connection is plugged in correctly and
make sure it is in good working order.

• Clean the balance and the housing, using a slightly moistened cloth.

Simply apply any commercially available cleaning agent available for
domestic use.

• Do not use any agents for cleaning that may scratch the surface

mechanically. Among other things it is covered in a conductive coat that is
important for safe operation.

• The adjustable feet of the platform balance and the shock absorbers

are made of rubber. When they are inside the hazard area, you will
have to use a slightly moistened cloth for cleaning. Moistness
protects against electrostatic charges.

• Make sure that you do not damage the earth contact springs during

cleaning. They serve as a retainer of electrostatic charge on the
weighing pan. They should be approx. 5 mm above the shock
absorbers.

• Do not clean in the presence of explosive gasses.

• The use of concentrated acids, alkaline solutions or alcohols for

cleaning purposes is prohibited.

• The same applies to the use of compressed air and high pressure

cleaning equipment.