Technical examination

• Technical examination by expert staff is required every three

months.


The examination should include the following:

• Connection to earth quality: The quality of the fastening on the housing and the

resistance of the connection -

max. 100Ω.

• The quality of the connection between the platform and the display unit (flexible

steel tube connecting both subassemblies).

• Damage and loss of a connection is inacceptable.
• The electrical connection between the housing and the platform – the maximum

resistance between the weighing pan (put on) should be less than100Ω.

• Stability and levelling of the balance.