13 data output rs 232 c, 1 technical data, 2 remote control instructions – KERN & SOHN CPB 30K5DM User Manual

Page 46

Advertising
background image

13 Data output RS 232 C


The balance is typically equipped with a RS 232C interface. Weighing data can be
edited according to menu setting or automatically or by pressing via the interface.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between
the weighing balance and the printer.

• Use a suitable cable to connect the weighing balance to the interface of the

printer. Faultless operation requires an adequate KERN interface cable.

• Communication parameters (baud rate, bits and parity) of weighing balance

and printer have to match, see chap. 11.2, Menu block „F4 AU“

.

13.1 Technical data

Connection

D-Sub 9 poles jack

Pin 2 output

Pin 3 input

Pin 7 signal earth

Baud rate

Selectable 600/1200/2400/4800/9600

Parity

8 bits, no parity

bold printed = factory setting

13.2 Remote control instructions

The remote control commands are sent from the remote control unit to the balance
as ASCII code. After the balance having received the commands, it will send the
following data.

Take into account that the following remote control commands must be sent without
a subsequent CR LF.

T

Tare placed weighing vessel

Z

Zeroing

C

Delete

P

Pieces to send

S

Send stable value

W

Send instable value


CPB-N / CPB-DM-BA-e-1424

47

Advertising
This manual is related to the following products: