16/18 Manual

7


Fig 4

Use numeric keypad to input wavelength (e.g. 500nm).

Press

ENTER

to change the wavelength, and then blank; After blanking,

the screen displays as the left one of Fig 5.

Note:1. If you don’t want to change the wavelength, you can press RETURN

】.

..

.

2. the wavelength range is 190-1100, other figure is invalid.

START

RETURN

Fig. 5

Measurement

(Example: set wavelength in “Basic mode”)

Push the blank cuvette into the light path. Then press

ZERO

】.

..

.

Press

START/STOP

in the left one of Fig. 5 to go into testing

interface.(The right one of Fig 5)

Press

START/STOP

again in the right one of Fig. 5, the testing

result displays on the screen, repeat this step, the results will
display on the screen one by one.

Print test report

(For example: Print the report in “Basic mode”, the right one of Fig. 5)

Press

PRINT

in the right one of Fig.5 to print the report (Fig. 6)

Fig 6

Note: After printing, all the data displayed on the screen will disappear.
Before measurement

Make a blank reference solution by filling a clean cuvette (or test tube) half full with

distilled or de-ionized water or other specified solvent. Wipe the cuvette with tissue

to remove the fingerprints and droplets of liquid. Fit the blank cuvette into the 4-cell

linear changer and place the cuvette in the slot nearest you. For the UV16\18,

push the changer so that the cuvette is in the light path (Push the rod in). Close the

lid.

Basic Mode

0.000

Abs

500.0 nm

500.0nm 0.000Abs

No. WL Abs

1

500.0 0 .000

Basic Mode

NO. Abs. T%

1 0.000 100.0%

2 0.000 100.0%