Specifications: installation – LABEC V-18 User Manual

Page 6

Advertising
background image

16/18 Manual

3

Packing List

Description Quantity

Spectrophotometer ...............................................................

1

Mains Lead ...........................................................................

1

Cuvettes................................................................................

Set of 4, glass

..........................................................................

Set of 2, quartz

Manual ..................................................................................

1

Note: the qurtz cuvette will not be included in the visible instruments.

Specifications:

Installation:

1. After carefully unpacking the contents, check the materials with the packing list

(page 2) to ensure that you have received everything in good condition.

2. Place the instrument (Fig.1-3) in a suitable

location away from direct sunlight. In order to have the best performance from

Model

63(&752V16 63(&752V18 63(&752UV16 SPECTRO-UV18

Wavelength Range

320-1100nm

190-1100nm

Spectral Bandwidth

4nm

2nm

4nm

2nm

Optical System

Single Beam, Grating 1200lines/mm

Wavelength Accuracy

±0.5nm

Wavelength Repeatability

0.3nm

Wavelength Resolution

±0.1nm

Photometric Accuracy

±0.3%T

Photometric Repeatability

±0.2%T

Photometric Range

-0.3-3A, 0-200%T

Stray Light

0.05%T@360nm

0.05%T@220nm,340nm

Stability

±0.002A/h @500nm

Display

Graphic LCD (128X64 dots)

Keyboard

22 Membrane keypad

Photometric Mode

T, A, E

Detector

Si Photodiode

Sample Compartment

Standard 10mm path length cuvette

Accommodates 100mm path length cuvette with optional holder

Light Source

Tungsten lamp

Tungsten & Deuterium

Output

USB Port Parallel Port (printer)

Power Requirement

AC 220V/50Hz or 110V/60Hz

Dimensions (W x D x H)

470 x 370 x 180mm

Weight

12kg

14kg

Advertising
This manual is related to the following products:

V-16, UV