ThermoProbe TL1-A User Manual

Thermo probe, Report of test, Calibration data as found calibration data as left

Advertising
background image

Report No.:
Model:

TL1-A

1-11923

75 °F +/- 2º

MS

Calibration Data As Found

Calibration Data As Left

In

Tolerance

º F

º C

º F

º C

º F

º C

º F

º C

º F

º C

º F

º C

32

0

32.011

0.006

32.03

0.02

-0.02

-0.01

0.10

0.06

Yes

0.030

0.017

120

49

120.015

48.897

120.02

48.90

0.00

0.00

0.10

0.06

Yes

0.030

0.017

199

93

199.022

92.790

199.05

92.81

-0.03

-0.02

0.10

0.06

Yes

0.030

0.017

300

149

300.230

149.017

300.23

149.02

0.00

0.00

0.10

0.06

Yes

0.030

0.017

R0: 100.00

A:

B:

C:

Date:

Callendar Van Dusen Coefficients:

The results stated on this report relate only to the items specifically identified.

6/4/2013

This test report or calibration certificate shall not be reproduced except in full,

without written approval of the laboratory.

3.91400E-03

-6.03510E-07

-4.17100E-11

Tested temperatures and corrections are as follows:

Tolerance

Correction

Measurement

Uncertainty

2013-06-04

Unit SN:

- 1-11923

New Unit or no "As Found" data available

This device has been adjusted to read as closely as possible to actual temperature.

Reading of TL1

Actual Test Temp.

Nominal Value

Calibration Date: 6/4/2013

Ambient Temp:

Calibrated By:

This report is to certify that the instrument listed below
has been calibrated by ThermoProbe, Inc. to NIST
traceable criteria.

Report of Test

Traceability: This calibration is traceable to NIST through an unbroken chain of comparisons.
IsoTech calibrated the references shown above using transfer standards which in turn were
calibrated by their primary reference thermometer.

Uncertainty Statement: Uncertainties were computed using the concepts, methods and
techniques of the ISO Guide to the Expression of Uncertainty in Measurement (the GUM). The
calculated uncertainty is an expanded uncertainty (k=2). It does not consider errors due to
possible damage to the TL1 from shipping, temperature drift, or thermal hysteresis effect. To
maintain the accuracy of the TL1, users should take care to protect it during shipping, avoid using
it to measure temperatures significantly above the highest calibrated temperature, and have the
TL1 recalibrated annually.


Calibrator's Signature: ___________________________________________


Test Results Approved by: _______________________________________



Test Method: The calibration procedures used were ThermoProbe, Inc. Calibration Procedures based on ASTM E-

644-06. This probe was immersed in a constant temperature bath with a reference thermometer which determined

the actual test temperature. The readings were compared and correction factors for the probe were calculated. The
As Left readings reflect the TL1's readings after calibration

Report of Test

ThermoProbe

Bath

Fluid

Reference

Calibration Date

Next Calibration

Due

(-)20.0º C

(-)4.0º F

Fluke 7340

water/glycol

TL1R-6068

3/27/2013

3/27/2014

0.0º C

32.0º F

Fluke 7340

water/glycol

TL2-0016

3/27/2013

3/27/2014

48.9º C

120.0º F

PolyScience 8101

mineral oil

TL2 - 0008

7/20/2012

7/20/2013

92.8º C

199.0º F

Fluke 6330

mineral oil

Fluke 1502A - ASP WSP500

3/27/2013

3/27/2014

149.0º C

300.2º F

Fluke 6330

Silicon oil

Fluke 1502A - ASP WSP500

7/20/2012

7/20/2013

290.0º C

554.0º F

Fluke 9172

metrology well Fluke 1502A - ASP WSP500

7/20/2012

7/20/2013

Nominal Temp