ThermoProbe TL2 User Manual

Report of test, Calibration data as found calibration data as left

Advertising
background image

Report No.:

- 0024

Model:

TL2

TL2-0024
AB3E29

MS

Calibration Data As Found

Calibration Data As Left

In

Tolerance

Calibration Date: 10/31/2012

Ambient Temp: 23ºC +/- 2ºC

Calibrated By:

New or Repaired Unit: No As Found data available

2012-10-31

CHANNEL 2

Unit SN:

Probe SN:

This device has been adjusted to read as closely as possible to actual temperature.
Tested temperatures and corrections are as follows:

Measurement

Uncertainty

Tolerance

Correction

Reading of TL2

Actual Test

Temp.

Nominal Value

Report of Test

This report is to certify that the instrument listed
below has been calibrated by ThermoProbe, Inc. to
NIST traceable criteria.

Traceability: This calibration is traceable to NIST through an unbroken chain of comparisons.
IsoTech calibrated the references shown above using transfer standards which in turn were
calibrated by their primary reference thermometer.

Uncertainty Statement: Uncertainties were computed using the concepts, methods and

Test Method:

The calibration procedures used were ThermoProbe, Inc. Calibration Procedures based on ASTM E-

644-06. This probe was immersed in a constant temperature bath with a reference thermometer which
determined the actual test temperature. The readings were compared and correction factors for the probe were
calculated. The As Left readings reflect the TL2's readings after calibration

Report of Test

Bath

Fluid

Reference

Calibration Date

Next Calibration

Due

(-)20.0º C

(-)4.0º F

Fluke 7340

water/glycol

TL1R - 6068

4/4/2012

4/4/2013

0.0º C

32.0º F

Fluke 7340

water/glycol

TL2 - 0016

4/4/2012

4/4/2013

48.9º C

120.0º F

PolyScience 8101

mineral oil

TL2-0008

7/19/2012

7/19/2013

92.8º C

199.0º F

Fluke 6330

mineral oil

Fluke 1502A - ASP WSP500

4/4/2012

4/4/2013

149.0º C

300.2º F

Fluke 6330

silicon oil

Fluke 1502A - ASP WSP500

7/19/2012

7/19/2013

290.0º C

554.0º F

Fluke 6330

silicon oil

Fluke 1502A - ASP WSP500

7/19/2012

7/19/2013

Nominal Temp

Tolerance

º F

º C

º F

º C

º F

º C

º F

º C

º F

º C

º F

º C

-4

-20

-4.009

-20.005

-3.986

-19.992

-0.023

-0.013

0.050

0.030

Yes

0.030

0.017

32

0

32.007

0.004

32.015

0.008

-0.008

-0.004

0.050

0.030

Yes

0.030

0.017

120

49

119.997

48.887

119.998

48.888

-0.001

-0.001

0.050

0.030

Yes

0.030

0.017

199

93

199.000

92.778

199.003

92.779

-0.003

-0.002

0.050

0.030

Yes

0.030

0.017

300

149

300.195 148.997

300.201 149.001

-0.006

-0.003

0.050

0.030

Yes

0.030

0.017

R0: 199.98

A:

B:

C:

Date:

without written approval of the laboratory.

3.91513E-03

-5.94034E-07

0.00000E+00

Uncertainty

10/31/2012

The results stated on this report relate only to the items specifically identified.
This test report or calibration certificate shall not be reproduced except in full,

Temp.

Callendar Van Dusen Coefficients:

Uncertainty Statement: Uncertainties were computed using the concepts, methods and
techniques of the ISO Guide to the Expression of Uncertainty in Measurement (the GUM). The
calculated uncertainty is an expanded uncertainty (k=2). It does not consider errors due to
possible damage to the TL2 from shipping, temperature drift, or thermal hysteresis effect. To
maintain the accuracy of the TL2, users should take care to protect it during shipping, avoid
using it to measure temperatures significantly above the highest calibrated temperature, and
have the TL2 recalibrated annually.

Calibrator's Signature: ___________________________________________

Test Results Approved by: _______________________________________

Title: _________________________________________________________

Advertising