The pumps are designed for

ambient temperatures

during operation between +50°F and +104°F (+10°C

and +40°C). Periodically check maximum tempera-

tures if installing the pump in a cabinet or a housing.

Make sure ventilation is adequate to maintain recom-

mended operating temperature. Install an external au-

tomatic ventilation system if necessary. If pumping hot

process gases, make sure that the maximum permitted

gas inlet temperature is not exceeded. The maximum

permitted gas inlet temperature depends on several

parameters like inlet pressure and ambient tempera-

ture (see “Technical data”, pg. 24).

+

Particles and dust must not enter the pump.

+

Do not pump liquids.

• Ensure that the pump is chemically resistant to the

pumped substances prior to operation.

Use the equipment

only as intended, that is, for genera-

tion of vacuum in vessels designed for that purpose. Any

other use will automatically invalidate all warranty and li-

ability claims. Remain aware of safety and risks.

Setting up and installing the equipment

Equipment must be connected only to a

suitable elec-

trical supply and a suitable ground point. As such,

the plug must be plugged into an outlet that is properly

grounded. Failure to connect the motor to ground may

result in deadly electrical shock.

The supply cable may be fitted with a molded Europe-

an IEC plug or a plug suitable for your local electrical

supply. The cable contains wires color coded as fol-

lows: green or green and yellow: ground; blue or white:

neutral; brown or black: hot.

+

Due to the high compression ratio, the pump may gen-

erate overpressure at the outlet. Check pressure com-

NOTICE