During operation – VACUUBRAND MZ 1C User Manual

Page 35

Advertising
background image

page 35 of 102

1.

Disconnect the electrical power cord.

2. Use a screw driver to adjust the voltage se-

lection switch to the supply voltage:

”115/120” corresponds to 100-120 V and

”230/240” corresponds to 200-230 V.

pump

switched off

pump

switched on

I

O

ON/OFF switch

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.

max.

104°F

(40°C)

+

Maximum ambient temperature: 104 °F

(40 °C)

Check the maximum temperatures, if install-

ing the pump in a cabinet or a housing.

Make sure ventilation is adequate, especial-

ly if the ambient temperature is elevated.

Operation with silencer at the outlet:
+

Operating the pump at a high inlet pressure

or pumping dusty gases for a long time may

cause clogging of the silencer. Check the

silencer regularly and replace if necessary,

or install a hose nozzle instead.