Read commands ”cvc 2000 – VACUUBRAND CVC 3000 User Manual

Page 62

Advertising
background image

page 62 of 80

Command

Operation

Response

Description

IN_PV_1

current pressure

XXXX mbar/

Torr/hPa

unit according to preselections

IN_PV_2

current frequency

XX.X Hz

pump speed

IN_CFG

device set

preselections

0XXXX

1XXXX

2XXXX

3XXXX

X0XXX

X1XXX

XX0XX

XX1XX

VACUU•LAN

continuous pumping

vacuum control without automatic

vacuum control with automatic

no coolant valve

coolant valve

no venting valve

venting valve

Read commands CVC 2000

Pin assignment RS-232 C

2: RxD

5: Mass

9: +5V (Bluetooth)

3: TxD

7: RTS

4: DTR

8: CTS

In remote mode (”Remote On”, with the ”PC symbol” in the display) all

keys at the controller are inoperable.

To return to the manual operation of the controller, set the control-

ler to ”Remote off” in menu configuration: Switch off the controller.

Then switch the controller back on, and press the selection knob

within 2s.

A maximum of ten commands per second is possible.

Read commands and commands ”REMOTE”, ”CVC”, and ”STORE”

can always be sent. The sending of other write commands is only pos-

sible, if ”Remote on” is selected.

The commands have to be written in capital letters.

Command and parameter have to be separated by a blank.

The string is terminated with <CR> or <LF> or <CR><LF>.

The response of the controller is always terminated with <CR><LF>.

Numerical values and parameters can be written without leading ze-

ros.

The response of the controller always includes leading zeros.