Program function – VACUUBRAND CVC 3000 E User Manual

Page 38

Advertising
background image

page 38 of 75

Program function

Permits ten programs to be defined and stored, each with up to

ten program steps with preset values for vacuum and time.

+

Edit:

Use to define the preset values for the process run:

Time: Defines either the process runtime for each program step to

reach a preset vacuum level or, if programming a ”Step”, the runtime

after having achieved the vacuum level. The total process runtime is

shown in the base line. Attention: A preset runtime of 99:59:59 hours

in the final program step will cause the process to run endlessly. Ter-

minate the process by pressing the ”STOP” key.

Vacuum: Vacuum level to be attained.

Venting valve: Operating a venting valve to reach a preset vacuum

level.

”Step”: ”Step” causes pump down as fast as possible to the preset

vacuum level. As soon as the vacuum level is reached the time meter

starts running.

+

Open: Choose program (Programs 0 - 9).

+

Store: This command stores an edited program or the program of the

last process to one of the storage spaces 0 - 9.

+

Hysteresis: The ”Hysteresis” is the control bandwidth of the two-point

control. A too small hysteresis will lead to frequent switching of the

valve or the pump. A too large hysteresis will lead to imprecise con-

trol. Suggested ”Hysteresis” values are stored in the controller (setup

”Auto”) and are adapted automatically to the preset pressure. The hys-

teresis can be adapted at any time.

The ”Hysteresis” is adjustable in a range of 1-225 Torr (1-300 mbar) or

can be set to ”Auto”.

+

Delay: Delay” determines the time the coolant valve remains open af-

ter the process has been stopped. Determines also the time the pump

(only with VMS module and in-line valve) remains running after the

process has been stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to