Gauge head VSP 3000 (accessory)
Read the manual of the gauge head VSP 3000 carefully before in-
stalling or operating the equipment. Observe the instructions con-
tained in that manual.
The display and measuring range of the CVC 3000 E with gauge head
VSP 3000 is from 7.5 x 10
2
Torr to 1 x 10
-3
Torr (1 x 10
3
mbar to 1 x 10
-3
mbar). The pressure reading is given in exponential notation.
+
Attention: Maximum pressure reading: 7.5x10
2
Torr
(1x10
3
mbar). Pressure values above 7.5x10
2
Torr
(1x10
3
mbar) are not displayed.
Risk of unnoticed
overpressure. Risk of bursting!
Note: The active pressure transducer
(VSP 3000) is not displayed in the sta-
tus line of the controller display!
When operating the CVC 3000 E
with the external VSP 3000 pressure
transducer,
vacuum control is not
possible in the ”Auto mode” and
” VACUULAN” functions.
1.0
x
10
3
mbar
”Pump down” function:
”Minimum” is adjustable in a range of 1x10
-3
Torr to 7.5x10
2
Torr (1x10
-3
mbar to 1x10
3
mbar) or can be set to ”Off”.
”VAC control” function:
”Set vacuum” (factory set 7.5x10
2
Torr), ”Hysteresis” and ”Maximum” are
adjustable in a range of 7.5x10
2
Torr to 1x10
-3
Torr (1x10
3
mbar to 1x10
-3
mbar).