Gauge head VSP 3000 (accessory)

Read the manual of the gauge head VSP 3000 carefully before in-

stalling or operating the equipment. Observe the instructions con-

tained in that manual.

The display and measuring range of the CVC 3000 E with gauge head

VSP 3000 is from 7.5 x 10

2

Torr to 1 x 10

-3

Torr (1 x 10

3

mbar to 1 x 10

-3

mbar). The pressure reading is given in exponential notation.

+

Attention: Maximum pressure reading: 7.5x10

2

Torr

(1x10

3

mbar). Pressure values above 7.5x10

2

Torr

(1x10

3

mbar) are not displayed.

Risk of unnoticed

overpressure. Risk of bursting!

Note: The active pressure transducer

(VSP 3000) is not displayed in the sta-

tus line of the controller display!

When operating the CVC 3000 E

with the external VSP 3000 pressure

transducer,

vacuum control is not

possible in the ”Auto mode” and

” VACUULAN” functions.

1.0

x

10

3

mbar

”Pump down” function:

Minimum” is adjustable in a range of 1x10

-3

Torr to 7.5x10

2

Torr (1x10

-3

mbar to 1x10

3

mbar) or can be set to ”Off”.

”VAC control” function:

Set vacuum” (factory set 7.5x10

2

Torr), ”Hysteresis” and ”Maximum” are

adjustable in a range of 7.5x10

2

Torr to 1x10

-3

Torr (1x10

3

mbar to 1x10

-3

mbar).