Adjustment under vacuum, 750 torr, 0 torr – VACUUBRAND CVC 3000 E User Manual

Page 51

Advertising
background image

page 51 of 75

Adjustment at atmospheric pressure

An adjustment at atmospheric pressure is only

possible in the upper pressure range.

Vent the measurement connection of the VSP 3000 pressure trans-

ducer. Make sure that the measurement connection of the VSP 3000

is at atmospheric pressure. Allow the pressure transducer a period of

minimum 20 minutes to preheat at this pressure.

In ”Configuration” menu, select program ”Adjustment” at the controller.

The reading is automatically adjusted to ”750 Torr” (”1000 mbar”). This

value is fix and can not be changed.

Press the selection knob to confirm.

750 Torr

Allow the pressure transducer a period of minimum 20

minutes to preheat. Adjustment is not possible during

the preheat time (20 minutes after connecting the volt-

age supply).

---- Torr is displayed, if adjustment is not possible in

the actual pressure range.

Adjustment under vacuum

An adjustment under vacuum is only possible

in the lower pressure range.

Evacuate the measurement connection of the VSP 3000 pressure

transducer preferably below 7.5x10

-4

Torr (1x10

-3

mbar), e.g., by ap-

plying a high vacuum pump. Allow the pressure transducer a period of

minimum 20 minutes to preheat at this pressure.

In ”Configuration” menu, select program ”Adjustment” at the controller.

+

The reading is automatically adjusted to ”zero”. This value is fix and

can not be changed.

Press the selection knob to confirm.

0 Torr