VACUUBRAND CVC 3000 E User Manual

Page 63

Advertising
background image

page 63 of 75

Command

Operation

Parameter Description

OUT-SP_1

set vacuum

XXXX

unit according to preselection; see re-

spective function for parameter range

OUT_SP_V

set vacuum with

venting

XXXX

unit according to preselection; see re-

spective function for parameter range

OUT_SP_2

speed

XXX

speed in % or ”HI”

OUT_SP_3

start-up pressure

XXXX

unit according to preselection; see re-

spective function for parameter range

OUT_SP_4

delay

XX:XX

hh:mm (hours:minutes)

OUT_SP_5

switch-off

pressure

XXXX

unit according to preselection; see re-

spective function for parameter range

OUT_SP_6

switch-off time

XX:XX

hh:mm (hours:minutes)

OUT_SP_PL

open program

X

program 0......9

OUT_SP_PS

store program

X

program 0......9

OUT_SP_P1y

time

XX:XX:XX

+XX:XX:XX

total runtime until program step y

(0.....9) or time for program step y

(0.....9) (additive)

OUT_SP_P2y

pressure

XXXX

pressure at program step y (0......9)

unit according to preselection

OUT_SP_P3y

venting valve

0

1

no venting valve in program step y (0...9)

venting valve in program step y (0......9)

OUT_SP_P4y

Step

0

1

no ”Step” in program step y (0.....9)

”Step” in program step y (0......9)

OUT_SP_P5y

Auto

0

1

2

no ”Auto” in program step y (0.....9)

”Auto ” in program step y (0......9)

”Auto ” in program step y (0......9)

START

started

STOP

0

1

2

Stop and delete fault

Stop

Stop and adopt the set vacuum

REMOTE*

0

1

Remote off

Remote on

ECHO**

0

1

Echo off

Echo on,

write command with return value

CVC

2

3

CVC 2000 commands

CVC 3000 commands***

OUT_VENT

0

1

2

venting valve closed

venting valve open

venting until atmospheric pressure

(788 Torr (1050 mbar) at maximum)

STORE

store settings permanently,

if ”ECHO = 1” after realization

Advertising