Prevent any part of the human body from coming into

contact with vacuum.

+

Attention: At pressures above approximately 795 Torr

(1060 mbar) the pressure reading becomes incorrect

due to saturation of the pressure transducer. The dis-

play flashes. Release pressure immediately!

Risk of

bursting!

• You must take suitable precautions to prevent any dan-

gerous situation from arising if the controller switches

the in-line valve or a coolant valve, switches a vacuum

pump (in combination with a VMS module), or opens a

venting valve.

Attention: If the controller is set to Autostart, the pro-

cess will start immediately after a power failure with-

out pressing any further key. It is your responsibility to

ensure that automatic start-up of the system will not

lead to any dangerous condition. Provide appropriate

safety measures. Check prior to starting the process

whether the option ”Autostart” (menu: configuration) is

enabled.

Electronic equipment is never 100% fail-safe. This may

lead to an ill-defined status of the equipment or of other

connected devices. Provide appropriate protective mea-

sures to allow for the possibility of failure and malfunction.

The protective measures must also allow for the require-

ments of the respective application.

Maintenance and repair

Switch off the controller. Disconnect the power

supply and wait five seconds before starting mainte-

nance to allow the capacitors to discharge.

Note: The equipment may be contaminated with chem-

icals, which have been processed during operation.