VACUUBRAND CVC 3000 detect User Manual

Page 50

Advertising
background image

page 50 of 74

This screen-shot shows the factory-set values.

+

Autostart: If ”Autostart” is set to ”On” the controller restarts a running

process automatically after a mains failure. If this is unwanted, set

Autostart” to ”Off”.

Attention: If Autostart” is preselected, the process starts

immediately after power failure without pressing any fur-

ther key. It is the user’s responsibility to ensure that no

dangerous status of the system due to the automatic start-

up can occur and to provide appropriate safety measures.

If necessary, the user has to check

prior to starting the

process if the option ”Autostart” is enabled.

+

Defaults: If ”Defaults” is set to ”Load”, the controller is reset to factory

set values. All stored programs and parameters are deleted.

Configuration
Adjustment 760 Torr

RS-232...

Sensors...

Display...

Autostart

Off

Defaults Cancel

- - - - - - - Back - - - - - - -