Fault

Possible cause

Remedy

❑ No display.

➨ Power supply not

plugged in (wall plug)?

✔ Plug in power supply.

➨ CVC 3000 detect con-

troller switched off?

✔ Switch on controller.

➨ VACUU • BUS cable

of power supply not

plugged in at control-

ler?

✔ Plug in VACUU • BUS

cable at controller.

➨ Other than above men-

tioned causes?

✔ Contact local distributor.

❑ Display disappears. ➨ Too much load (e.g.,

valves) connected?

✔ Check current draw of the

connected devices.

➨ Short circuit at connect-

ed valves?

✔ Replace valves.

➨ Short circuit at the

RS 232 plug?

✔ Check plug and cable.

➨ Other than above men-

tioned causes?

✔ Contact local distributor.

❑ Pressure reading

incorrect.

➨ Pressure transducer

decalibrated?

✔ Readjust CVC 3000 detect

or external gauge head.

➨ Humidity in the pres-

sure transducer?

✔ Let the pressure trans-

ducer dry, e.g., by pump-

ing. Readjust if neces-

sary. Determine and

eliminate the cause for

humidity.

➨ Pressure transducer

contaminated?

✔ See ”Cleaning the pres-

sure transducer”.

➨ Other than above men-

tioned causes?

✔ Contact local distributor.

❑ Digital pressure

reading is flash-

ing, display shows

”0.0”.

➨ Pressure transducer

not correctly adjusted

under vacuum?

✔ Adjust CVC 3000 detect

or external gauge head

correctly.

❑ No digital pressure

reading.

➨ Pressure transducer

defective?

✔ Contact local distributor.

❑ Digital pressure

reading is flashing,

one blip*.

➨ Overpressure at the

pressure transducer,

pressure > 795 Torr

(1060 mbar)?

✔ Release pressure imme-

diately (risk of bursting).