VACUUBRAND CVC 3000 detect User Manual

Page 67

Advertising
background image

page 67 of 74

* only if ”Sound” ”On” is selected

** VMS: Vacuum management system to switch non-VARIO pumps

Fault

Possible cause

Remedy

❑ ”Vac control” func-

tion: Control stops,

”arrow up” is flash-

ing.

➨ Preset maximum pres-

sure exceeded?

✔ Confirm by pressing

START/STOP key.

Change maximum pres-

sure value if necessary.

❑ ”Pump down” func-

tion: Control stops,

”arrow down” is

flashing.

➨ Pressure below preset

minimum pressure?

✔ Confirm by pressing

START/STOP key.

Change minimum pres-

sure value if necessary.

❑ No function is dis-

played. No menu

available.

➨ No controllable device

is connected (valve,

VMS, NT VARIO /

VARIO-B pump)?

✔ Connect devices or use

controller as measuring

device.

❑ Controller does

not respond when

pressing keys (ex-

cept ON/OFF).

PC symbol is dis-

played.

➨ Controller in remote

mode?

✔ Control CVC 3000 detect

via interface or switch off

remote mode.

❑ Controller does

not respond when

operating any keys.

No change after

switching off/on.

✔ Contact local distributor.

Note: All error messages comprising a flashing warning triangle have to

be confirmed (deleted) by pressing the START/STOP key.

Solve the problem (see ”Remedy”), then press the START/STOP key to

delete the error message.